Startup/Shutdown
Typical state (System interface is on, Chillers are on):
Open command shell (2nd button down on left in CrysalisPro)
Type 'script' (no quotes) and hit return
Navigate to C:\Xcalibur\macro and run 'turn_on_xray.mac'
Wait a while (
Sign into the log book
CCD Detector information from CCD button in CrysalisPro
X-ray information from Xray button in CrysalisPro
Cryojet information from Cryo button in CrysalisPro
Chiller flows from ODBench on CCD command line
Click start on Xray Chiller &Â Ccd Chiller tabs
Run experiment (see other pages for details)
Shut down: run the 'turn_off_xray.mac' script and leave instrument as is
Ground up state (Nothing is on):
Turn on water supply (green valve handle should be opened enough to give 17.5 psi on pressure gauge)
Turn on System Interface (Mains on system interface - top of electronics cabinet)
Start the X-ray and CCD chillers (Mains buttons at bottom of electronics cabinet)
Turn on X-ray Generator (Power on - Nova X-ray generator - middle of electronics cabinet)
Go to above